Test Strips
If Control Solution results are out of range, your Bionime BGMS may not be working properly. Repeat
the Quality Control Test. If your Control Solution results outside the range still exist, do not use the
BGMS to test your blood glucose and contact your Bionime Customer service.
Quality Control Test
PRECAUTION
Each time you open a new bottle of Control Solution, write
the discard date on the label. Control Solution is good for 3
months after opening the bottle, or until the expiration date
printed on the label, whichever comes first.
The your Control Solution results are out of the range :
Y
dure correctly.
Malfunction of the meter or the test strip.
possible reasons
- our Control Solution has expired or has been opened 3 months ago.
- Your test strip has expired.
- You left the cap of the tst strip vial opened or the control solution off for a long time.
- You didn't perform the test proce
-
Quality Control Test
Please use Control Solution tested with GM550 BGMS under Control Solution Mode. If the
test result is within the Control Solution Range printed on the strip vial label, the
BGMS passes Quality Control Test. That means your BGMS is working correctly.
TM
Rightest
TM
Rightest GM550
TM
Rightest GM550
Example of Control Solution Range
printed on your test strip vial label.
Control Solution Range:
When should a Quality Control Test be performed ?
- Whenever you want to check if your system is working properly or not
- Whenever you want to practice testing and check correct procedure.
- To prepare for your initial blood glucose test.
TM
- To check the Rightest Blood Glucose Test Strip GS550 when you open a new vial of strips.
TM
- To check your Rightest Blood Glucose Meter GM550 after it has been dropped, damaged or
exposed to liquids.
- When you suspect that your test results are inaccurate, or if your test results are not consistent
with the way you feel.
.
Example of Expiration Date:
Exp. Date 2012-06
Control Solution Range:
83-113 mg/dL
4.6-6.3 mmol/L