43
42
Sampling Error
Please don't apply the blood drop to the sample entry of the strip before the
meter displays . If you do so, the meter will display " " and " "
accompanied by beeps (if volume is turned on). Please discard the test strip and
repeat the test with a new teststrip.
" "
Meter Malfunction
If the meter can not be started, please follow the steps below:
1. Open the battery cover, remove the batteries.
2. Wait for 5 minutes and insert the batteries as described “Meter activation and Battery change”
on page 16~17.
The meter should be working normally after finishing above steps. If not, please contact your local
Bionime customer service.
Temperature Error
In order to get accurate test result, perform testing between 10~40℃ ( 50~104℉ ).
1. When the ambient temperature is 0~9℃ ( 32~48℉) or 41~50℃ (106~122℉)
the " " warning symbol will be blinking (1a), you still can do the test but the test
result is only for reference, because the test result under these ranges of
temperature might not be correct. In the memory screen this value will be flagged
with symbol (1b). Repeat the test at an area with temperature between
operating range. ( 10~40℃ or 50~104℉ )
2. When you move from the area with temperature outside the operating range of
test strip to another area with temperature inside operating range of test strip,
please wait for 30 minutes before you do the test.
3. When the temperature is below 0℃ ( 32℉ ) or over 50℃ ( 122℉ ), Meter can not
do the test and the " " symbol will blink with thermograph symbol displaying
on screen in this condition. Please move the meter to environment with
temperature between 10~40℃ ( 50~104℉ ) and repeat the test after 30 minutes.
" "
1a
1b
3
Code Error
It means the strip is not inserted right or to the proper position. Please reinsert the
strip following the instruction. If Code Error repeats again, it means you use wrong
strip or strip not suitable for your region test. After you make sure you use the right
strip and the strip is inserted right, but error screen still appears, please contact
your local Bionime Customer Service.
Error messages and Trouble shootingError messages and Trouble shooting