View Window
This window is yellow before applying
blood sample. It gradually turns red
when filled with blood.
Hand Bar
Grip here to insert
test strip into meter.
Sample Entry
µL
Apply a drop of blood or control solution here.
The test request is only 1.0 of blood.
Electrode Contacts
Sensing signal output terminals.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol
up and toward meter.
TM TM
The Rightest GM550 Meter must only be used with Rightest GS550 Test Strips and
TM
Rightest Control Solution.
The application of other Test Strips or control solutions can lead to incorrect results.
PRECAUTION
- Re-cap the test strip vial immediately after taking out a test strip.
- Do not reuse test strips.
- Do not use expired test strips.
- Record the date of opening a new test strip vial for the first time. Discard the vial of test
strips after 3 month from opening.
- Store the test strips in a cool (4-30) and dry location (< 90% relative humidity). Do not
expose to direct sunlight or heat.
- For detailed information, please refer to the Test Strips Package Insert.
TM
Rightest GS550
TM
- If the Rightest GM550 meters and strips are exposed to a high temperature difference,
please wait 30 minutes before measurement.
GS550 Test Strip GS550 Test Strip